Local Operation 4

Tutorial

gld08.bmp

Figure 4-6. Main Menu - Control Mode

7.Press ENTER to confirm the mode change. The upper left corner will change to CONTROL, and the pressure will start moving towards the set-point.

8.After the pressure is stabilized, assure that the word Step is highlighted in the Step/Jog [F5]. The Step/Jog [F5] key toggles between Step and Jog, to switch between these modes, press the function key 5 [F5]. In addition to using the keypad to enter a pressure set-point, you can also use the step function. Using the rotary knob, rotate the knob clock-wise to increase the set-point, or counter clock-wise to decrease the controller set-point. The scratchpad will be updated with a new set-point. The step size defaults to 10% of full scale. This step size is user defined and can be changed in the Menu Setup User menu.

9.Press ENTER to accept the new set-point. The Calibrator will move to the new set-point.

10.Press Measure key which is located in the key-pad. The Calibrator will change to Measure mode. No confirmation is necessary to leave Control mode.

Selecting Language

The RUSKA 7350 Calibrator can operate in a variety of different languages. To select a different language, press the Mode key two times MODE, MODE. The current language will be highlighted. Use the rotary knob to highlight a new language and hit the ENTER key to select that language.

Selecting Mode of Operation

The RUSKA 7350 Calibrator normally operates in Absolute mode but also supports Tare mode for calibration of Gauge mode sensors. As an option for increasing accuracy by limiting range, the RUSKA 7350 Calibrator may be ordered with Triple Ranging.

Tare Mode

Press the Mode key then press Tare [F3] to turn on Tare mode. The current pressure reading will be stored as the tare offset and the pressure reading will be zeroed. Press Tare [F3] again to turn off Tare mode.

4-5

Page 29
Image 29
Fluke 7350 specifications Selecting Language, Selecting Mode of Operation, Tare Mode