ANSI/IEEE 488.2-1987 Command Summary

*CLS

 

Clear Status

*ESE?

 

Event Status Enable Query

*ESE

<number>

Event Status Enable

*ESR?

 

Event Status Register

*IDN?

 

Identification

*OPC?

 

Operation Complete Query (Returns 1)

*OPC

 

Operation Complete

*RST

 

Reset

*SRE?

 

Service Request Enable Query

*SRE

<number>

Service Request Enable

*STB?

 

Status Byte Query

*TST?

 

Self-Test Query

*WAI

 

Wait (No operation)

SCPI Command Summary

The current value associated with a SCPI command may be read by appending a question mark to the command. For example CALC:LIM:UPP? will return the current upper pressure limit.

MEASure

 

 

[:PRESsure]?

returns current pressure reading

:POSITION?

returns pump position (%)

CALCulate

 

 

:LIMit

 

 

:LOWer

<number>

get/set low pressure limit

:SLEW

<number>

get/set slew rate limit

:UPPer

<number>

get/set high pressure limit

:VENT <number>

get/set auto-vent limit

:TARE

 

 

:VALue <number>

get/set tare value

:STATe ON/OFF

set tare using current pressure

CALibration

 

 

[:PRESsure]

 

:Data

 

 

:POINts ?

number of calibration constants

:VALue<n>?

returns cal constants label, value

:VALue<n> <number>

set calibration constant

:CALibration

 

:POINts?

number of calibration points

:VALue?

Nominal calibration point

:VALue<n> <number>

perform calibration point

:ZERO

 

performs zero calibration

:VALUE <number>

sets zero value

:INITiate

enter zero calibration mode

:INITiate?

returns status for cal, press, temp, ref

:RUN

start zero calibration

:STOP

abort zero calibration

:POSition

 

 

:DATA <number>, <number>

sets C0, C1 position

:DATA?

 

reads C0, C1

:VALue <number>

sets first calibration point

:VALue 2 <number>

sets second calibration point

5-4

Page 56
Image 56