Remote Operation

5

 

 

Device Messages

:MODE?

 

calibration edit enabled?

 

:MODE ON¦OFF¦1¦0

 

enable calibration edit

 

DISP

 

(Cal. button required)

 

 

 

 

:ENABle ON¦OFF¦1¦0

turns front panel display on/off

 

:TEXT <string>

 

displays message on front panel

 

:BGRaph <number>

 

sets bar graph maximum

 

OUTPut

 

 

 

:STATe ON¦OFF¦1¦0

 

off=MEASure, on=CONTrol

 

:STATe?

 

returns 0=Measure or 1=Control

 

:MODE MEASure¦CONTrol¦VENT

sets mode

 

:MODE?

 

returns mode string

 

PROGram

 

 

 

:CATalog?

 

returns list of defined programs

 

[SELected]

 

 

 

:DEFine <program block>

Define program

 

 

 

press1, toler1, dwell1, max1, press2,

 

 

 

toler2, ..

 

:DEFine?

 

read program definition

 

:DELete

 

 

 

[:SELected]

 

deletes current program

 

:ALL

 

deletes all programs

 

:NAME <program name>

select current program

 

:STATe RUNPAUSeSTOPCONTinue

set program state

 

:STATe?

`

read program state

 

:CONFigure

 

 

 

:RECall

 

restore saved configuration

 

:SAVE

 

save current configuration

 

SENSE

 

 

 

[:PRESSure]

 

 

 

[:RESolution] <number>

set pressure display resolution

 

:AUTO <boolean>ONCE

return to default resolution

 

:MODE?

 

returns ABSOLUTE or TARE

 

:RANGE

 

 

 

[:UPPer]

 

returns RUSKA 7350 full scale value in

 

units

 

 

 

:LOWer

 

returns 0

 

:REFerence

 

 

 

[:HEIGht] <number>

set head height

 

:SGRavity <numbers>

set specific gravity

 

[SOURCE]

 

 

 

[:PRESsure]

 

 

 

[:LEVel]

 

 

 

[:IMMediate]

 

 

 

[:AMPLitude] <number>

sets pressure setpoint

 

[:AMPLitude]?

read pressure setpoint

 

:MODE FIXedLIST

set source parameter set

 

:TOLerance <number>

specifies output tolerance

 

:SLEW <number>

 

set slew rate

 

:CONTrol <number>

set control band

 

5-5

Page 57
Image 57
Fluke 7350 specifications Mode?