Local Operation 4

Menu

gld23.bmp

Figure 4-21. Menu Program Config - Menu

5.Press PREVIOUS until the Main Menu appears.

6.Press Setup [F2] Limits [F1]. Change the desired parameters using the normal procedures.

7.Return to the Main Menu by pressing PREVIOUS three times.

8.Press Menu [F6] Program [F4].

9.Use the rotary knob to highlight the name of the program.

10.Press Config [F3] Save [F1]. The configuration of the Calibrator will be changed to the current settings and stored in the program’s configuration.

Running a Program

1.Programs are run from the Program Menu. From the Main Menu (press PREVIOUS until the Main Menu appears), press Menu [F6] Program [F4].

2.Use the rotary knob to highlight the name of the program.

3.Press Run [F1]. The program run screen will appear. “Stop” will be highlighted, showing that the program is not currently running.

gld24.bmp

Figure 4-22. Menu Program Run - Menu

4-23

Page 47
Image 47