Description and Corrective Action

 

 

-286

Program Runtime Error. An error occurred while running the program. Usually the

 

setpoint is out of range.

 

 

-313

Calibration Data Lost. The calibration data has been lost and the unit must be

 

recalibrated.

 

 

-315

Configuration Data Lost. The configuration data has been lost. Check all parameters

 

to be sure that they are correct.

 

 

-330

Self Test failed. See display for more information.

 

 

-350

Queue Overflow. The error queue was full and messages were lost.

 

 

-400

Query Error. A read request was received when there was nothing to read.

 

 

500

Controller Malfunction. Internal control failure.

 

 

501

High Limit Exceeded. The pressure was greater than the high limit.

 

 

502

Low Limit Exceeded. The pressure was less than the low limit.

 

 

503

Slew Limit Exceeded. The pressure changed faster than the slew limit allowed.

 

 

521

Pressure Overrange. The pressure reading is outside the range of the RUSKA 7350.

 

 

538

Auto vent. The Vent Limit was exceeded and the instrument entered Vent Mode.

 

 

543

Supply Pressure. The instrument is attempting to increase pressure and the

 

pressure is not responding. Check main pressure supply, shop air supply, and for

 

leaks.

 

 

600

Factory Data Lost. Internal factory constants have been lost. Contact Fluke for more

 

information.

 

 

602

Model Changed. Firmware has been downloaded that is not correct. Download the

 

correct firmware.

 

 

706

Pump position failure. The position sensor is not detecting movement in the pump

 

after the pump has been commanded to move.

 

 

707

Range Switch. The current range can only be changed when the pressure reading is

 

below 100 psi (700 kPa).

 

 