Fluke 7350 specifications Summary of Error Messages

Models: 7350

1 78
Download 78 pages 53.35 Kb
Page 77
Image 77

Appendix B

Summary of Error Messages

Summary of Error Messages

Negative error numbers are from the Standard Commands for Programmable Instruments (Version 1991.0).

 

Table B-1. Summary of Error Messages

 

 

Value

Description and Corrective Action

 

 

0

No Error.

 

 

-103

Invalid Separator. Check punctuation in the SCPI command.

 

 

-104

Data Type. The type of parameter data is incorrect. Check for numeric versus string

 

data.

 

 

-109

Missing Parameter. No valid parameter was found for the SCPI command.

 

 

-110

Command Header. The command name is not valid.

 

 

-113

Command Unknown. The command specified does not exist.

 

 

-114

Header Suffix. The numeric suffix for the command name is out of range.

 

 

-201

Remote Required. Remote command conflicted with local interface. Return local

 

interface to main Menu.

 

 

-221

Settings Conflict. The command could not be executed due to the current state of

 

the Calibrator. Some commands cannot be executed while a program, self-test, or

 

calibration is in progress.

 

 

-222

Out of Range. The value is not within the valid range. For pressures, check your

 

high and low limits.

 

 

-281

Cannot create program. Program memory is full.

 

 

-282

Illegal Program Name. The name specified is not valid or does not exist.

 

 

-284

Program Currently Running. The command cannot be executed while a program is

 

running.

 

 

-285

Program Syntax Error. The syntax of the program definition is not correct.

 

 

B-1

Page 77
Image 77
Fluke 7350 specifications Summary of Error Messages