RUSKA 7350
Users Manual
Note
If the actual pressure is outside of the tolerance for the requested
Step 3
1.To begin Step 3, use the calibration standard to Apply the
2.When the Measured pressure stabilizes, use the RUSKA 7350’s numeric keypad and OK to enter the actual pressure reported by the calibration standard. Do not enter the Measured pressure reported by the RUSKA 7350. If necessary, use the CLEAR key to correct a mistake in the edit field. If the actual pressure applied is acceptable, the Calibration, Step 4, screen will appear.
Note
If the actual pressure applied is outside of the tolerance for the requested
Step 4
1.To begin Step 4, use the calibration standard to Apply the
2.When the Measured pressure stabilizes, use the RUSKA 7350’s numeric keypad and OK to enter the actual pressure applied by the calibration standard. Do not enter the Measured pressure reported by the RUSKA 7350. If necessary, use the CLEAR key to correct a mistake in the edit field. If the actual pressure applied is acceptable, the Calibration complete screen will appear.
Storing the Coefficients
Note
In addition to saving the calibration coefficients to the RUSKA 7350’s memory, the user is advised to separately record the calibration coefficients and store this “backup” in a safe place.
Step 5
Calibration is complete. To exit the calibration procedure without storing the calibration coefficients in memory, press CANCEL. To store the calibration coefficients in memory, select DONE.
Step 6
Press PREV. to return to the main screen.
Once the calibration procedure is complete, the user is advised to record several pressure readings. If there are any significant errors at these points, then an error was probably made in generating one of the calibration pressures, and the calibration procedure should be repeated.