RUSKA 7350

Users Manual

:LIST

 

:PRESsure <number> [,<number>]

set list of pressure values

:POINts?

returns number of points defined

:DWELl <number> [,<number>]

specifies dwell times

:POINts?

returns number of dwell times

:TOLerance <number> [,<number>]

specifies tolerances

:POINts?

returns number of tolerances

:DIRection UPDOWN

direction to go through list

:COUNt <number>

number of times to go through list

STATus

 

:OPERation

 

[:EVENT]?

read/clear operation event register

:CONDition?

read operation condition register

:ENABle <number>

set operation enable mask

:QUEStionable

 

[:EVENT]?

read/clear questionable event register

:CONDition?

read questionable condition register

:ENABle <number>

set questionable enable mask

:PRESet

reset condition flags

SYSTem

 

:DATE <year>,<month>,<day>

set system date

:ERRor?

returns <error#,”descr;info”>

 

or 0,”No Error”

:KLOCk ONOFF10

lock keyboard

:TIME <hour>,<minute>,<second>

set system time

:VERSion?

returns 1991.0

:LANGuage “6000””SCPI”

set interface protocol to 6000 or SCPI

:PRESet

reset system

TEST

 

:ELECtronic?

perform electronic self-test

UNIT

 

:DEFine<n> <name>,<number>

define a unit

:LENGth MMIN

set length units for head height

[:PRESsure] <unit name>

set pressure units

Example SCPI Commands

To request the current pressure reading, all of the following commands are equivalent:

:MEASURE:PRESSURE?

:measure:pressure?

:MeAsUrE:pReSsUrE?

:meas:pres?

:measure?

:meas?

MEAS?

To set the control pressure setpoint to 50, all of the following commands are equivalent:

SOURCE:PRESSURE:LEVEL:IMMEDIATE:AMPLITUDE 50 SOUR:PRES:LEV:IMM:AMPL 50.0

PRESSURE +50 PRES 50

5-6

Page 58
Image 58
Fluke 7350 specifications List, Example Scpi Commands