Fluke 7350 specifications Menu Test Sweep Test

Models: 7350

1 78
Download 78 pages 53.35 Kb
Page 48
Image 48

RUSKA 7350

Users Manual

4.Press Run [F2]. The configuration of the Calibrator stored with the program is restored, the pressure set-point is set to the pressure value in the first step, and the Calibrator is placed in Control mode. “Run” will now be highlighted and the program will proceed through its steps.

5.To pause the program, press Pause [F3]. “Pause” will now be highlighted and the Calibrator will continue controlling to the current set-point. The Calibrator will maintain control of the current set-point until further instructions from the operator. Press Continue [F4] to resume the program.

6.To stop the program, press Stop [F5]. The program will stop running but the Calibrator will continue controlling to the current set-point.

Menu Test

The Menu Test section of the system is used to perform a number of the diagnostic and tuning functions that are available with the RUSKA 7350.

gld25.bmp

Figure 4-23. Menu Test - Menu

Menu Test — Sweep Test

The sweep function can be used to automatically exercise the elastic sensing element of the device under test prior to performing a calibration. For the Calibrator to perform the sweep test, input the high and low set-point pressures, the pressure control tolerance, the dwell time at set-point, and the number of cycles to perform.

1.Ensure that the units, limits, and control parameters are set to their desired values.

2.The sweep test is entered from the Sweep Test Menu. From the Main Menu, (press PREVIOUS until the Main Menu appears), press Menu [F6], then Test [F5], and then Sweep [F2].

3.Enter the high and low pressure points, the control tolerance, dwell time, and the number of cycles.

4-24

Page 48