Offset Temperature Drift: The zero offset voltage will drift less than ± 1 count at the display over the operating range of 10°C to 35°C.

High Performance Battery Charger

Battery Type and Capacity: 6V (3 cell) Lead Acid (5 AH) provide 8 hours of continuous operation.

Recharge Time: Fast recharge time on the order of 2 - 3 hours even during operation.

Battery Life Span: On the order of 8-10 years due to multiple level charging circuit (2000 charge/discharge cycles typically).

Low Battery Annunciator: Displayed when 30 minutes or less of operation remains.

Environment

Operating: 0°C to 35°C, up to 70% relative humidity.

Storage: -25°C to +65°C.

Mechanical

Dimensions: 21.6 cm wide x 8.9 cm high x 26 cm deep (10 in. x 9.4 in. x 4 in.)

Weight: 5 kg (11 lbs.)

1.6.3 Interface Cable - Detector to Display

Part Number: 37783-X where X denotes cable length in meters.

Standard Length: 15 m and 30 m.

Maximum Total Length: 60 m ( 197 ft.).

Ion Chamber Signal Cable Type: Five channels of RG-174 low noise coaxial.

Leakage: Less than 0.5 pA over the full operating temperature range.

Connector Type: Display interface end has five AMP male coaxial connectors and three AMP male single conductor connectors housed in a single DB-37 type housing. Detector interface end has a five AMP female coaxial connectors and three AMP female single conductor connectors housed in a single DB-37 type housing. Multiple cables may be interconnected to form other lengths.

Minimum Conduit Diameter: 7.6 cm (3 in.) without disassembly. Cable may be run through smaller opening by using special tool for disassembly. Consult factory for further details.

Weight: 2.25 kg (5 lbs.) at 15 m (weight is approximately proportional to cable length for other lengths).

* Specifications subject to change without notice.

1.7 Unpacking and Inspection

The 90100 TRACKER System was carefully inspected before shipment. Upon receiving the 90100 TRACKER System, carefully unpack all items from the shipping carton and check for any obvious signs of physical damage that might have occurred during shipment. Report any damage to the shipping agent at once. Retain the original packing material in case reshipment becomes necessary. The items in Figure 1-3 are included with every 90100 TRACKER System.