where Measurement is the value on the 35360A TRACKER Display, and ICCF and DUCF are defined on page 60 and page 59 respectively.

Enter the value calculated in Step 2 into the Display Calibration Items screen of the Customization software. Be sure to enter it in the correct column and row. For example, if you are setting the charge gain for the Center chamber, enter the value calculated in Step 1 in the third column and first row.

If you wish to use the Front Panel Gain Factor and the Air Density Correction Factor, be sure to enable them after the calibration is finished.

5.8 Usage Examples

CHANGING THE POWER DOWN INTERVAL

1.Place the cursor on the Power Down Interval Item.

2.Press F2. The field clears.

3.Enter 10 and press Enter.

4.The software indicates that it is downloading. When the Item reappears, the download was successful. The number that appears is the value returned from the 35360A TRACKER Display.

CHANGING THE PRESSURE UNITS

1.Place the cursor on the Pressure Units Item.

2.Press F2. The function key area indicates the keys used to make a choice.

3.Press the spacebar or use the up and down arrow keys to change the units.

4.Select millibars and press Escape. The original units remain and the Customization software returns to Selection mode.

5.Repeat steps 2 and 3. Select millibars and press Enter.

6.The software indicates that it is downloading. When the Item reappears, the download was successful.

CHANGING THE DISPLAY UNITS CONVERSION FACTOR

CAUTION

This example changes the current value of a factory-set conversion factor. Make a note of the value BEFORE changing it so that it may be set to this value at the end of the example.

1.Place the cursor on the Display Units Conversion Factor Item.

2.Press F2. The field clears.

3.Enter 0.765 and press Enter.

4.The software indicates that it is downloading. When the Item reappears, the download was successful. The number that appears is the value returned from the 35360A TRACKER Display.

5.Press F2.

6.Enter the original value and press Enter.

7.The software restores the original value in the instrument.