Fluke 90100 manual Model 35360A Tracker Display, Measurements Mode, Display Mode

Models: 90100

1 60
Download 60 pages 32.95 Kb
Page 9
Image 9

General Information

Specifications 1

1.6.2 Model 35360A TRACKER Display

Measurements Mode

The measurement mode is user-selected from the front panel prior to the start of the exposure.

Dose (rad): Electrometer is operated in integrating mode with capacitive feedback.

Dose Rate (rad/min): Electrometer is operated in non-integrating mode with resistive feedback.

Display Mode

The display mode is user-selected from the front panel and may be selected before, during, or following the exposure.

Absolute: All five ion chamber measurement values are displayed in absolute dose measurement units (typically rads or rad/min) as defined by the dose measurement unit's description.

Percent-of-Center: Only the center ion chamber measurement value is displayed in absolute dose measurement units (typically rad or rad/min) as defined by the dose measurement unit's description.

Four peripheral ion chamber measurement values (top, bottom, left, and right) are displayed as a percentage of the center ion chamber measurement value.

Electrometer / A-D Converter

Nominal Full Scale Output: 3 volts.

A/D Converter Resolution: 16 bits unipolar.

Nominal Full Scale and Sensitivity in Electrical Units:

A-D Dose

A-D Dose

A-D Rate

A-D Rate

Full Scale

Sensitivity

Full Scale

Sensitivity

1000.0 nC ± 2%

0.015 nC/count

1000.0 nC/min ± 2%

0.015 (nC/min)/count

 

 

 

 

An overrange indication will be displayed when the input current or charge have exceeded the full-scale values. The full-scale value of any unit is guaranteed not to differ from nominal values by more than ± 2%.

Long Term Stability: Expected to be less than 0.25% of reading per year.

Non-linearity: Less than ± 0.4% of range.

Display Resolution: 0.003% of range (4-3/4 digits) for both dose and dose rate. Display resolution is limited by the following relationship for all display units;

1 A-D Count 1 Display Count < 10 A-D Counts.

This relationship insures that the resolution of the measurement display values is as high as possible without exceeding the inherent resolving capability of the

A-D converter.

Gain Temperature Stability: Less than ± 0.4% of reading over operating range of 10°C to 35°C.

Measurement Noise: Charge and Current measurement noise are less than ± 1 count at the display exclusive of source noise effects. Typical effects from a Co60 source are on the order of ± 0.2 rad/min

at 100 rad/min. Source noise effects during dose rate measurements on a linear accelerator may be excessive.

Measurement Accuracy: Fluke Biomedical factory calibration will provide a basic Charge (Current) measurement accuracy of ±1% of reading ± 0.3 nC (± 0.3 nC /min) exclusive of long-term drift and temperature effects. Similar results may be obtained in the field using a suitable calibration source such as the Keithley Instruments, Inc. Model 263 Calibrator/Source and the included Customization software.

Current Leakage: Less than 1 pA (typically 1 rad in 5 minutes)

Zero Offset Voltage: Factory trimmed to 0 counts at the display.

1-5

Page 9
Image 9
Fluke 90100 manual Model 35360A Tracker Display, Measurements Mode, Display Mode, Electrometer / A-D Converter