Fluke 90100 manual Basic Operating Quantities

Models: 90100

1 60
Download 60 pages 32.95 Kb
Page 35
Image 35

 

 

 

 

Theory of Operation

3

 

 

 

 

Basic Operating Quantities

3.6 Basic Operating Quantities

 

 

 

Table 3-3. Basic Operation Quantities

 

 

 

 

 

 

 

 

 

 

 

Quantity

Description

Units

Comments

 

 

 

 

 

 

 

 

 

 

 

 

 

Electrical current delivered to readout in

 

 

 

I

Current

nC/Min

response to radiation incident upon

 

 

 

 

 

 

detector.

 

 

 

 

 

 

Electrical charge delivered to readout in

 

 

 

Q

Charge

nC

response to radiation incident upon

 

 

 

 

 

 

detector.

 

 

 

 

 

 

Electrometer output level returned by

 

 

 

ADC

A-D Counts

Counts

A-D Converter during dose or dose rate

 

 

 

measurement.

 

 

 

 

 

 

 

 

 

 

 

 

Electrometer output level returned by

 

 

 

 

 

 

A-D Converter while operating in the

 

 

 

 

 

 

dose rate measurement mode and in

 

 

 

ADZ

A-D Zero

Counts

the absence of radiation. A-D Zero

 

 

 

levels are field programmable via

 

 

 

 

 

 

 

 

 

 

 

 

Customization Software.

 

 

 

 

 

 

Overall electrical charge gain of the

 

 

 

 

 

 

electrometer / A-D combination when

 

 

 

 

 

 

measuring in the dose mode. Charge

 

 

 

QG

Charge Gain

nC / Count

gain values are field programmable via

 

 

 

 

 

 

Customization Software.

 

 

 

 

 

 

Overall electrical current gain of the

 

 

 

 

 

 

electrometer and A-D combination

 

 

 

 

 

 

when measuring in the dose rate mode.

 

 

 

IG

Current Gain

(nC/Min) / Count

Current gain values are field

 

 

 

programmable via Customization

 

 

 

 

 

 

 

 

 

 

 

 

Software.

 

 

 

 

 

 

Specific R/nC values are supplied by

 

 

 

 

Ion Chamber

Ion Chamber

factory for each of the five ion chamber

 

 

 

 

channels. Values are field

 

 

 

ICCF

Calibration

Calibration Units

 

 

 

programmable via Customization

 

 

 

 

Factor

/ nC

 

 

 

 

Software.

 

 

 

 

 

 

 

 

 

DUCF

Display Units

Display Units

Factory default of 0.879 rad/R is

 

 

 

Conversion

/ Radiation Calibration

supplied. Value is field programmable

 

 

 

 

Factor

Units

via Customization Software.

 

 

 

 

Air Density

 

Calculated by instrument and depends

 

 

 

ADCF

Correction

None

on front panel Temperature and

 

 

 

 

Factor

 

Pressure settings.

 

 

 

FPGF

Front Panel

None

User specified via front panel gain

 

 

 

Gain Factor

setting.

 

 

 

 

 

 

 

3-7

Page 35
Image 35
Fluke 90100 manual Basic Operating Quantities