Maintenance & Service

Troubleshooting 4

35360A TRACKER Display will not turn on (display will not light)

 

NOTE

 

35360A TRACKER Display's

 

 

 

VFD does not illuminate when

 

 

 

When attempting to turn the instrument on,

POWER key is pressed.

firmly press the POWER key and hold

 

for approximately one second. Always

 

allow several seconds between retries.

 

WARNING

The Model 35360A TRACKER Display and 35300A Detector contain no internal operator serviceable

Disassembly of either unit could result in electrical h k

Always turn the 35360A TRACKER Display's power off

prior to connecting or disconnecting the interface bl

Failure to do so may result in an electrical h k

When powering the 35360A TRACKER Display from

an AC line, always use a 3-wire grounding type line cord h

as the one supplied with the 90100 TRACKER System. In addition, only connect the 35360A TRACKER Display to

AC power outlet employing a third wire safety d

Yes

No

Operating from AC power source?

Yes

 

 

Connect 35360A TRACKER Display to an AC power source and press the POWER key.

35360A TRACKER Display

No

 

VFD illuminates?

 

Commence operation from AC line power or charge batteries.

Verify voltage present at AC line

receptical by connecting and powering some other device.

Yes

AC line voltage present?

No

 

 

Check AC line fuses at rear panel of 35360A TRACKER Display.

No

Replace 35360A TRACKER

Display's AC line fuses. Refer to Section 5.3.

Fuses OK?

Yes

 

Return 35360A TRACKER Display to Keithley for repair.

Figure 4-1. No Display Flowchart