Fluke 90100 manual Operation Positioning the Model 353300A Detector in the Beam

Models: 90100

1 60
Download 60 pages 32.95 Kb
Page 17
Image 17

Operation

2

Positioning the Model 353300A Detector in the Beam

Gantry arm

Treatment head

Model 35300A

Gantry Detector

Treatment

To Model 35360A

table

TRACKER Display

Gantry

Treatment table

Cable

To Model 35360A

TRACKER Display

Figure 2-4. Setup for 0-Degree Gantry Angle (NOTE: Build-up not shown for clarity)

For those who have purchased the optional Build-Up Peg and Retainer Kit, gantry angles other than 0 degrees may be used. For gantry angles of 90 or 270 degrees, place the 35300A Detector on its side so the label faces the head. Orient the connector away from the gantry. This placement is shown in Figure 2-5.

WARNING

Operating the 35300A Detector on edge without build-up pegs constitutes a tip-over hazard and could result in personal injury.

2-5