Fluke 90100 manual Nominal Number of build-up Solid Water

Models: 90100

1 60
Download 60 pages 32.95 Kb
Page 58
Image 58

Model 90100

Operators Manual

Recommended build-up for photons is given in Table A-1; recommended build-up for electrons is given in Table A-2. Table A-2 takes into account the 0.56 cm water-equivalent inherent in the 35300A Detector. Cobalt-60 radiation does not require any added build-up.

Table A-1. Recommended Amount of Build-up to Use With the Model 35300A Detector to Attain Photons dmax for Linear Accelerators

Nominal

Number of build-up

"Solid Water"

Accelerator Potential,

plates

thickness (cm)

MeV

(37051C-1)

 

2

0

0

3

1

0.5

4

2

0.5

5

2

1.0

6

3

1.0

7

3

1.5

8

3

1.5

9

4

1.5

10

4

2.0

11

4

2.0

12

4

2.0

13

5

2.0

14

5

2.5

15

5

2.5

16

6

2.5

17

6

3.0

18

6

3.0

19

6

3.0

20

7

3.0

21

7

3.5

22

7

3.5

23

7

3.5

24

8

4.0

25

8

4.0

 

 

 

A-2

Page 58
Image 58
Fluke 90100 manual Nominal Number of build-up Solid Water