Read Time-Out

Use this selection to set a time-out (in milliseconds) of the scanner’s trigger when using serial commands to trigger the scanner. Once the scanner has timed out, you can activate the scanner either by pressing the trigger or using a serial trigger command. After scanning the Read Time-Outbar code, set the time-out duration (from 0-300,000 milliseconds) by scanning digits from the inside back cover, then scanning Save. Default = 300,000.

Read Time-Out

Manual Trigger, Low Power

The scanner powers down until the trigger is pulled. When the trigger is pulled, the scanner powers up and operates until there is no triggering for the time set with the Low Power Time-Outbar code below. There is a delay of up to one second in operation when the scanner is first triggered, but there is no delay when operating in low power time-out mode.

Manual Trigger, Low Power

Note: Manual Trigger, Low Power cannot be used with keyboard wedge applications.

Low Power Time-Out Timer

Scan the Low Power Time-Out bar code to change the time-out duration (in seconds). Then scan the time-out duration (from 0-300 seconds) from the inside back cover, and Save. Default = 120 seconds.

If there are no trigger pulls during the “low power time-out timer” interval, the scanner goes in low power mode. Whenever the trigger is enabled, the “low power time-out timer” is reset.

Low Power Time-Out

Automatic Trigger

The scanner scans continuously at full power with illumination fully on.

Note: If the Automatic Trigger selection is enabled, the aimer beam option is disabled.

Automatic Trigger

3 - 4