3 - 4 IT5600/IT5800 User’s Guide

Read Time-Out

Use this selection to set a time-out (in milliseconds) of the scanner’s trigger when
using serial commands to trigger the scanner. Once the scanner has timed out,
you can activate the scanner either by pressing the trigger or using a serial
trigger command. After scanning the Read Time-Out bar code, set the time-out
duration (from 0-300,000 milliseconds) by scanning digits from the inside back
cover, then scanning Save.
Default = 300,000.

Manual Trigger, Low Power

The scanner powers down until the trigger is pulled. When the trigger is pulled,
the scanner powers up and operates until there is no triggering for the time set
with the Low Power Time-Out bar code below. There is a delay of up to one
second in operation when the scanner is first triggered, but there is no delay
when operating in low power time-out mode.
Note: Manual Trigger, Low Power cannot be used with keyboard wedge
applications.

Low Power Time-Out Timer

Scan the Low Power Time-Out bar code to change the time-out duration (in
seconds). Then scan the time-out duration (from 0-300 seconds) from the inside
back cover, and Save.
Default = 120 seconds.
If there are no trigger pulls during the “low power time-out timer” interval, the
scanner goes in low power mode. Whenever the trigger is enabled, the “low
power time-out timer” is reset.
Automatic Trigger
The scanner scans continuously at full power with illumination fully on.
Note: If the Automatic Trigger selection is enabled, the aimer beam option is
disabled.
Read Time-Out
Manual Trigger, Low Power
Low Power Time-Out
Automatic Trigger