System Programming

Network Menu

10.Press Enter to confirm or Esc to cancel. If confirmed, the controller calibrates the sensor’s Zero. This takes only a few moments and the display returns to the default calibration screen.

Note: Never calibrate any unit’s Zero with ambient air. Always use Nitrogen (N2) at the calibration port to calibrate the Zero.

11.To calibrate the device, scroll to Set Span** and change the span gas calibration value using this procedure;

Calib 001 : 301D2

Status :

Normal

Set Zero

 

 

 

Set Span

 

 

2.60 ppm

 

 

a.Using the right arrow, move the cursor to xxx PPM (span value field). Press Enter to select the field (it is editable when flashing).

b.Use the up or down arrows to increase or decrease the value, press Enter to validate the new value.

c.Move the cursor back to Set Span and press Enter to start the calibration.

Calib 001 : 301D2

Status : In calib...

Set Zero

 

Set Span

2.60 ppm

The device Span is being calibrated. The screen will display the device’s status as “In calib...” until the calibration is complete.

**When selecting Set Span, make sure that the device has been supplied with the appropriate calibration gas before and during the calibration process.

64

301C User Manual

M-510324

 

Honeywell

7/10

Page 66
Image 66
Honeywell 301c user manual Calib 001 301D2 Status Normal Set Zero Set Span 60 ppm