14Occlusion Effect Test

14.1Occlussion effect measurement

1)Refer to 10: Real-Ear Measurement Setup to prepare the system for use and for proper positioning of the client and the probe tube.

2)Insert the earmold or custom hearing instrument into the ear, being careful not to advance the probe tube further into the ear canal. A lubricant applied to the earmold or custom instrument shell in the vicinity of the probe tube will make insertion easier and prevent slit leaks around the tube.

3)Turn the hearing aid OFF.

4)Press <Tests>, then highlight and <PICK> [Occlusion].

5)Highlight and <PICK> [Start test] to start the test. [Start test] changes to [Stop Test].

6)While the client vocalizes “ee” in a normal voice, highlight and <PICK> [Stop test] to stop the test and freeze the screen. A red occlusion bar indicates an occlusion problem, a yellow bar indicates a possible problem and a green bar indicates no problem.

7)Increase venting and repeat the test until a green bar is obtained.

FastFacts 14.1: Occlusion effect measurement

0611

RM500SL User’s Guide Version 2.8