HP RM500SL manual 20.11REM speaker overdriven, FastFacts 20.12 Barcode scanner malfunction

Models: RM500SL

1 112
Download 112 pages 21.85 Kb
Page 92
Image 92

20.11REM speaker overdriven

REM speaker overdriven!

This message indicates that the REM speaker is unable to produce the SPL required for a test. This is usually the result of a standing wave null at the REM reference microphone caused by reverberation in the environment. It warns of a situation that could cause significant measurement errors.

vMove the RM500SL and client away from walls and hard objects. Avoid placing the RM500SL under an overhanging shelf or cabinet.

vMove the RM500SL to the front edge of the table.

vMove the client slightly closer to or further from the REM speaker.

vPosition the test ear more directly in front of the REM speaker.

vApply sound absorbers to nearby reflective surfaces.

FastFacts 20.11: REM speaker overdriven

20.12Barcode scanner malfunction

If there is no red scanning beam when the trigger is pulled, tug on the cable where it enters the scanner. If it comes loose easily, push it back in until it clicks in place.

If the scanning beam stays on when the trigger is released or if the barcode scanner fails to read known good barcodes, try the following:

1)Unplug the scanner from the RM500SL, then reconnect it. If the problem persists…

2)Print this page on an external printer (see 2.7: External printer, auxiliary audio outputs) or from the electronic user’s guide on the RM500SL software CD (see 1.3: Electronic user’s guide).

3)Scan these two barcodes, starting with the top one. This will reset the scanner to its default settings.

FastFacts 20.12: Barcode scanner malfunction