18Speechmap Fitting Procedures

18.1Section overview

The procedures in this section assume a familiarity with 10: Real-Ear

Measurement Setup, 11: REM Screen Setup and 17: Speechmap ®.

This section covers the following topics:

18.2: Assessment data entry

18.7: Fitting to targets for mid-level speech

18.3: Assessment data choices

18.8: Open fittings in Speechmap

18.4:Assessment data choices - 2

18.9: FM fitting and verification

18.6:Fitting to targets for loud sounds

 

Related sections are 17: Speechmap ® and 19: Speechmap Technical Details

FastFacts 18.1: Speechmap fitting procedures

18.2 Assessment data entry

Threshold data and parameters may be entered by scanning a barcode on a previous RM500SL or Verifit printout or manually using a mouse, the keypad or the numeric keypad on an external keyboard. See 2.6: Mouse, keyboard, barcode scanner, and 3.2: Input device operation.

To enter data from a barcode see 3.3: Barcode data input and:

1)Press the barcode trigger and center the red line on the barcode from a distance of 6 - 8”. The scanner will beep and the Barcode Entry poster opens.

2)Highlight and <PICK> [Done] to accept the data indicated on the poster.

To enter assessment data manually:

1)Highlight and <PICK> [Audiometry].

2)Highlight and <PICK> an item to toggle its setting or generate a list box. Age can be changed only if Targets = [DSL child]. See 18.3: Assessment data choices - 1 and

18.4: Assessment data choices - 2.

3)Press <Continue> to proceed to data entry.

4)Enter HL threshold using a mouse,keypad or keyboard. Left click the mouse on the audiogram to enter a point. Click again to delete the point. On the keypad the arrow keys change frequency and level and <PICK> enters or deletes a threshold point. On a keyboard enter HL values on the numeric keypad.

5)Press <Continue> when all points have been entered.

6)Repeat the process for Bone conduction, UCL and RECD if [Enter] has been selected for any. If RECD is set to [Measure], see 15: RECD (real-ear to coupler difference).

FastFacts 18.2: Assessment data entry

0611

RM500SL User’s Guide Version 2.8