HP RM500SL manual FastFacts 18.5 Fitting to targets for soft speech

Models: RM500SL

1 112
Download 112 pages 21.85 Kb
Page 80
Image 80

18.5 Fitting to targets for soft speech

See 17.9: REM or S-REM mode, 17.6: Using Speechmap and 18.2: Assessment data entry for setup and audiometric data entry.

1)Select [Audiometry] to choose the target method and enter audiometric data.

2)Select REAR 1 and set stimulus to [Speech-std(1)] or [Speech-std(2)] and level to 50. Elongated + symbols show the target range for 50 dB speech LTASS.

3)Adjust frequency shaping and gain for soft sounds so that the middle curve (the LTASS) falls within the target range, especially between 500 - 4000 Hz. Press <Continue> to run the complete passage. Repeat as necessary.

FastFacts 18.5: Fitting to targets for soft speech

0611

RM500SL User’s Guide Version 2.8

Page 80

Page 80
Image 80
HP RM500SL manual FastFacts 18.5 Fitting to targets for soft speech