HP RM500SL manual FastFacts 16.3 Insertion gain in HL

Models: RM500SL

1 112
Download 112 pages 21.85 Kb
Page 65
Image 65

16.3 Insertion gain in HL

See 10: Real-Ear Measurement Setup and 11: REM Screen Setup

1)Press <Tests>, then highlight & <PICK> [Insertion gain] from the REM tests list.

2)Highlight and <PICK> [Instrument] to chose from BTE, ITE, ITC, CIC, Body. This changes the CORFIG used for 2cc targets.

3)Highlight and <PICK> fitting method to chose from FIG6, NAL-NL1, NAL-R,

POGO II, BERGER, LIBBY, USER, NONE.

An estimated aided threshold curve is calculated by shifting the unaided threshold curve upward by the REIG. This estimate assumes a) the unaided threshold is a sound-field threshold and b) a linear aid (i.e. REIG doesn’t depend on level).

The REUR curve uses the right-hand SPL scale and is shown for reference only.

FastFacts 16.3: Insertion gain in HL

0611

RM500SL User’s Guide Version 2.8

Page 65

Page 65
Image 65
HP RM500SL manual FastFacts 16.3 Insertion gain in HL