9.7Manual control procedure
Manual control displays the reference microphone SPL, the coupler microphone SPL, their difference in dB (gain) and coupler microphone signal distortion at any selected 1/12th octave frequency and at a selected input level from 40 to 90 dB (in 5 dB steps) and at 0 dB (off).
1)Press <Tests>, then highlight and <PICK> [Manual control].
2)Highlight and <PICK> the Level window, then highlight and <PICK> the desired stimulus level from the drop- down list.
3)Highlight and <PICK> the Frequency window, then highlight and <PICK> the desired frequency from the
4)Highlight and <PICK> [Start test] to start the test. [Start test] changes to [Stop Test]. To stop the measurement, highlight and <PICK> [Stop test]. You may change Level and Frequency without stopping the test.
FastFacts 9.7: Manual control procedure
9.8Sound level meter using manual control
When set for a stimulus level of 0 dB (off), the Manual control mode allows the RM500SL to be used as a sound level meter. Sound at the HIT reference microphone and the coupler microphone may be analyzed by 1/12th octave, A- weighted or
1)Press <Tests>, then highlight and <PICK> [Manual control].
2)Highlight and <PICK> the Level window, then highlight and <PICK> 0 dB stimulus level from the
3)Highlight and <PICK> the Filter window, then highlight and <PICK> 1/12th octave,
4)For 1/12th octave filter, highlight and <PICK> the Frequency window, then highlight and <PICK> the desired frequency from the
5)Highlight and <PICK> [Start test] to start the test. [Start test] changes to [Stop Test]. To stop the measurement, highlight and <PICK> [Stop test]. You may change Level and Frequency without stopping the test.
NOTE: A microphone extension cable
FastFacts 9.8: Sound level meter using manual control