1.5Notices for the European Community

Authorised Audioscan representative within the European Community:

PC Werth Ltd., Audiology House, 45 Nightingale Lane, London SW12 8SP UK

Phone: +44 (0)20 8772 2700 www.pcwerth.co.uk

Electromagnetic Compatability (EMC)

Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the following information:

vThe RM500SL should not be used adjacent to or stacked on other equipment. If this is necessary, its operation should be verified as normal in this configuration.

vPortable and mobile RF communications equipment can affect medical electrical equipment and may affect the performance of the RM500SL.

vThe RM500SL has not been tested for immunity to electromagnetic disturbances. Performance degradation due to such disturbances (including electrostatic discharge) is considered normal and acceptable.

FastFacts 1.5: Notices for the European Community

1.6Declaration of EMC compliance for the European Community

The compliances listed in the following table are met with the Audioscan SL-100 RECD transducer connected. The use of other accessories may result in increased emissions.

Guidance and manufacturer’s declaration - electromagnetic emissions

The RM500SL is intended for use in the electromagnetic environment specified below. The user of the RM500SL should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment - guidance

 

 

 

RF emissions

 

The RM500SL uses RF energy only for its

Group 1

internal function. Therefore, its RF emissions

CISPR 11

are very low and not likely to cause any

 

 

 

interference in nearby electronic equipment.

 

 

 

RF emissions

Class A

The RM500SL is suitable for use in all

CISPR 11

establishments other than domestic and those

 

directly connected to the public low-voltage

 

 

Harmonic emissions

Class A

power supply network that supplies buildings

IEC 61000-3-2

used for domestic purposes.

 

 

 

 

Voltage fluctuations/

Complies

 

flicker emissions

 

IEC 61000-3-3

 

 

 

 

 

FastFacts 1.6: Declaration of EMC compliance for the European Community

0611

RM500SL User’s Guide Version 2.8

Page 9

Page 9
Image 9
HP RM500SL manual Electromagnetic Compatability EMC, Cispr