19.4 Microphone location effects

Microphone Location Effects

Add to all stimuli in S-REM mode

 

 

Hearing instrument type

 

Frequency

BTE

 

ITE

ITC

CIC

 

BODY

200

0.5

 

0.3

0.0

0.9

 

3.0

250

0.5

 

0.5

0.3

0.8

 

3.0

315

0.8

 

0.8

0.3

0.7

 

3.0

400

1.1

 

1.0

0.7

0.5

 

3.0

500

1.2

 

1.8

0.0

0.4

 

4.0

630

1.1

 

2.0

0.1

0.4

 

3.0

800

0.9

 

2.0

0.4

0.4

 

2.0

1000

0.3

 

1.5

1.2

0.6

 

0.0

1250

0.6

 

0.3

-1.6

1.1

 

0.0

1600

2.5

 

-0.3

-1.9

2.0

 

-4.0

2000

4.1

 

3.8

2.1

3.3

 

-4.0

 

 

 

 

 

 

 

 

2500

3.5

 

5.0

4.8

5.0

 

-3.0

3150

2.8

 

3.3

3.5

6.9

 

-2.0

4000

3.7

 

4.3

6.4

8.3

 

0.0

5000

-1.2

 

4.3

6.6

7.6

 

0.0

6300

1.6

 

-0.4

-1.8

4.2

 

0.0

8000

3.3

 

1.0

-1.9

-5.6

 

0.0

Microphone location effects are added after the overall rms level has been set

FastFacts 19.4: Microphone location effects

0611

RM500SL User’s Guide Version 2.8

Page 86