8.3ANSI S3.22-2003 facts
ANSI
The most significant change from the 1996 version is the requirement for the hearing aid to be set in its most linear mode for the setting of the gain control to Reference Test Setting (changed from Reference Test Position) and for all tests except attack and release and
The OSPL90 curve, the HFA (or SPA) and peak value of this curve, the frequency of the peak, the HFA (or SPA)
These changes result in more consistent values for reference test gain, equivalent input noise and attack and release time. See 7.6: 1996 or 2003 ANSI standard, 8.2: ANSI
FastFacts 8.3: ANSI
8.4ANSI 1996 Linear and AGC tests
1)Press <Tests>, then highlight and <PICK> [Linear] or [AGC].
2)Highlight and <PICK> [Start test]; test setup instructions will be displayed.
3)Refer to 7.11: Coupling the hearing instrument, 7.12: Positioning the hearing
instrument and 2.5: Battery pill use and care to prepare the instrument for testing.
4)Turn the instrument ON.
5)Set the gain control (manual or programmable) to maximum with all other controls set to provide the widest frequency response, greatest gain and highest maximum output. AGC controls and/or programmable parameters should be set as designated by the manufacturer.
6)Press <Continue> to run the test or press <Cancel> to abort the test. Note that, if required, the RM500SL will pause during the test to request that the volume control be adjusted to the Reference Test Position (RTP). Open the HIT chamber if necessary and, without moving the hearing instrument, adjust the user gain control until the large vertical arrow on the screen prompt disappears. Press <Continue> to resume the test.
If you are unsure which test sequence to run, try [AGC] first. If attack and release times are all near zero, the instrument is probably linear.
FastFacts 8.4: ANSI 1996 Linear and AGC tests