16Insertion Gain

16.1Section overview

Insertion gain is the difference between aided and unaided ear canal SPL (REIG = REAG - REUG). It inherently assumes a sound-field audiogram. For non-linear hearing aids, REIG is stimulus-dependent and the dependency is specific to each hearing aid. Insertion gain fitting methods ignore these facts. Speechmap is recommended for fitting all non-linear hearing instruments.

Refer to 10: Real-Ear Measurement Setup and 11: REM Screen Setup for terminology and information on preparing the system for use and for proper positioning of the client and the probe tube.

This section covers the following topics:

16.2: Insertion gain in SPL

16.6: REAR measurement procedure

16.3: Insertion gain in HL

16.7:SII calculation in Insertion gain

16.4: Audiometric data entry

 

16.5: REUR measurement procedure

 

FastFacts 16.1: Insertion gain

16.2 Insertion gain in SPL

See 10: Real-Ear Measurement Setup and 11: REM Screen Setup

1)Press <Tests>, then highlight & <PICK> [Insertion gain] from the REM tests list.

2)Highlight and <PICK> [Instrument] to chose from BTE, ITE, ITC, CIC, Body. This changes the CORFIG used for 2cc targets.

3)Highlight and <PICK> fitting method to chose from FIG6, NAL-NL1, NAL-R, POGO II, Berger, Libby, User, None

FastFacts 16.2: Insertion gain in SPL

0611

RM500SL User’s Guide Version 2.8

Page 64

Page 64
Image 64