CONTRACTURE FOOTPLATE OPTION

PROCEDURE 11

CONTRACTURE FOOTPLATE OPTION

 

 

 

ANGLE ADJUSTMENT OF THE CONTRACTURE

FOOTPLATE (FIGURE 6)

 

Flat Screws

1. Loosen, but DO NOT, remove the two

Contracture

 

Assembly

 

(2) flat screws, washers and locknuts that

 

 

Contracture

secure contracture footplate to the

 

 

Footplate

footplate hinge.

 

 

 

2. Tilt the contracture footplate UP/

Footplate

 

DOWN to achieve desired angle.

Hinge

 

3. Retighten the two (2) flat screws,

NOTE: Mounting

 

washers and locknuts that secures

Washers

hardware removed

the contracture foot plate to the

for clarity only.

Lock Nuts

footplate hinge.

 

4. If necessary, repeat STEPS 1-3 on

NOTE: Bilateral contracture footplate shown

remaining contracture footplate.

for clarity.

 

 

FIGURE 6 - ANGLE ADJUSTMENT

 

OF THE CONTRACTURE

 

FOOTPLATE

CONTRACTURE FOOTPLATE DEPTH ADJUSTMENT

(FIGURE 7)

BILATERAL CONTRACTURE

NOTE: Observe the angle of the contracture

 

Flat Screws

 

 

footplate for reinstallation.

 

 

1. Perform one (1) of the following:

 

Mounting Hole

 

 

A.BILATERAL CONTRACTURE -

Remove the two (2) flat screws,

 

 

washers and locknuts that secure the

Contracture

 

contracture footplate to the footplate

 

Footplate

 

hinge.

 

 

 

B. CONTRACTURE PLATFORM -

 

Washers

Remove the four (4) flat screws that

Footplate Hinge

Lock Nuts

secure the contracture platform to

 

CONTRACTURE PLATFORM

the half clamps.

2. Move the contracture footplate to one (1)

 

of four (4) mounting hole.

Flat Screws

 

3. Perform one (1) of the following:

A.BILATERAL CONTRACTURE - Reinstall the two (2) flat screws, washers and locknuts that secure the contracture footplate to the footplate hinge.

B.CONTRACTURE PLATFORM - Reinstall the four (4) flat screws that secure the contracture platform to the half clamps. Tighten securely.

Contracture

Platform

FIGURE 7 - CONTRACTURE

FOOTPLATE DEPTH

ADJUSTMENT

Solara™

68

Part No. 1080556 Rev. I

Page 68
Image 68
Invacare 1080556 manual Angle Adjustment of the Contracture