PROCEDURE 8

LIMIT SWITCH

D.Only tighten the two (2) phillips screws, washers and locknuts that secure the actuator to the up- per limit switch bracket until the actuator does not move.

E.Repeat STEPS 1-3 until the wheelchair does not operate when the captain's seat back is at a 24o angle.

24o Back

114o

Recline Angle

5-3/4 inches

Magnetic Protractor - (Will

 

read 61o when back recline

61o

angle is at 114o relative to the

 

seat frame)

 

Upper Limit

 

Switch

 

Bracket

 

Phillips

 

Screws,

 

Washers

 

and Lock-

 

nuts

 

Actuator

 

5o Seat Angle

FIGURE 2 - ADJUSTING LIMIT SWITCH