Invacare Ranger II250 SERIES Replacing Batteries, M W D W H E E L C H A I R S, Procedure, Rear of

Models: Ranger IISTANDARD

1 84
Download 84 pages 3.62 Kb
Page 67
Image 67
Wheelchairs Equipped With MCC-Mark IV Joystick (Detail "A" of FIGURE 6)

MWD WHEELCHAIRS

PROCEDURE 14

NOTE: If performing the charging procedures indepen- dently, READ and CAREFULLY follow the individual in- structions for each charger (supplied or purchased).

NOTE: If charging instructions are not supplied, consult a qualified service technician for proper procedures.

Required Items:

TOOL

QUANTITY

COMMENTS

 

 

 

Battery Charger

1

Supplied

Extension Cord

1

Not Supplied

 

 

 

3-prong plug, 15 ampere current rating; industrial type

Wheelchairs Equipped With MCC-Mark IV Joystick (Detail "A" of FIGURE 6)

1.If necessary, remove the rear shroud. Refer to RE-

MOVING/INSTALLING SHROUDS - MWD WHEEL- CHAIRS in PROCEDURE 9 of this manual.

2.Attach the battery charger connector to the charger cable/battery harness.

3.Plug the charger’s AC power cord, or extension, into the grounded 120 VAC wall outlet.

4.Wait until charging is complete.

5.If necessary, install the rear shroud. Refer to RE-

MOVING/INSTALLING SHROUDS - MWD WHEEL- CHAIRS in PROCEDURE 9 of this manual.

Wheelchairs Equipped With Mark IV RII Joystick (Detail "B" of FIGURE 6)

1.Attach the battery charger connector to the charger port on the FRONT of the joystick.

2.Plug the charger’s AC power cord, or extension, into the grounded 120 VAC wall outlet.

3.Wait until charging is complete.

NOTE: Allow eight (8) hours for normal charging. Larger batteries (greater than 55 ampere-hours) or severely dis- charged batteries may require up to sixteen (16) hours to be properly charged and equalized.

NOTE: R2BASIC and R2STANDARD ONLY - If the charger operates for sixteen (16) hours and is unable to fully

charge the batteries, an internal timer turns the charger off and begins to fast blink the green light.

NOTE: If the batteries need to be charged more often or take longer to charge than normal, they may need to be replaced. Contact an Invacare dealer for service.

DETAIL "A"

Battery Charger Connector (on Side of Wheelchair)

WHEELCHAIRS

WITHOUT

SHROUDS

From Battery

Charger

 

Battery Charger

 

Connector (at

WHEELCHAIRS

Rear of

Wheelchair)

WITH SHROUDS

 

From Battery Charger

 

DETAIL "B"

 

 

 

 

Battery Charger

Three (3)

 

 

Pronged Plug
Front View of

From Battery

 

 

 

Joystick

Charger

 

 

 

 

 

 

 

 

 

 

 

 

 

Charger

Port

Joystick Wheelchairs Equipped With Mark IV RII Joystick (Detail "B" of FIGURE 6)

Battery DETAIL "A"

Boxes

Charger Port

FIGURE 6 - CHARGING BATTERIES

REPLACING BATTERIES

WARNING

The following procedure should be performed only by aqualified technician.

Most batteries are not sold with instructions. How- ever, warnings are frequently noted on the cell caps. Read them carefully, otherwise injury or damage can occur.

NOTE: Invacare recommends that both batteries be re- placed if one (1) battery is defective.