Nortec Industries GS Series manual Fault history, Tank Drain

Models: GS Series

1 78
Download 78 pages 6.36 Kb
Page 57
Image 57
Fault history

Y = total number of recorded services (up to 10 maximum).

STEAM HOURS XXX (DATE/TIME)

This field displays the steam hours and the date and time when the specific service (X) was performed.

SERVICE INTERVAL SERVICE HISTORY FAULT HISTORY

Tank Drain TANK DRAIN TANK PRE-CLEAN BURNERS

SHORT CYCLE DELAY

FAULT HISTORY X/Y

(TYPE OF FAULT)

(DATE/TIME)

Fault history

In this area the customer can view the history of recorded faults for the humidifier.

FAULT HISTORY X/Y

The fault history stores the last 54 system faults.

X = fault occurrence, where 1 is the most recent.

Y = total number of recorded faults (up to 54 maximum).

(type of fault) and (date/time)

A description of the fault type is displayed along with a date and time stamp when the fault occurred.

SERVICE INTERVAL

SERVICE HISTORY

FAULT HISTORY Manual background TANK DRAIN

TANK PRE-CLEAN BURNERS

SHORT CYCLE DELAY

THIS WILL DRAIN CONTENTS OF TANK Manual background

THIS WILL DRAIN CONTENTS OF TANK

CONTINUE?

Tank Drain

The tank drain feature, when activated, will activate the drain valve for a fixed period of time (depending on unit size).

To drain the tank press enter when CONTINUE is selected. During the tank drain cycle the display will indicate the tank is draining and the amount of time remaining on the cycle. To interrupt the cycle the unit must be switched off then on. Must enter service password to activate this function.

- 53 -

Page 57
Image 57
Nortec Industries GS Series manual Fault history, Tank Drain