ENTER READING CONFIGURATION MENU:
Press a1) Press a, if necessary, until CNFG prompt appears.
Press d2) Display advances to INPT Input Menu.
Press a3) Display advances to RDG Reading Configuration Menu.
Press d4) Display advances to DEC Decimal Point.
DECIMAL POINT SUBMENU:
Press d5) Display flashes previous selection for Decimal location.
Press b6) Scroll though the available selections and choose Decimal
location: FFFF, FFF.F, FF.FF or F.FFF
Press d7) Display shows STRD stored message momentarily only, if
changes were made, otherwise press ato advance to LOAD
Known/Unknown Loads Submenu.
Decimal Point is passive.
KNOWN/UNKNOWN LOADS SUBMENU:
Press d8) Display flashes previous selection of ENBL Enable or DSBL
Disable.
Press b9) Scroll though the available selection of ENBL or DSBL
(flashing).
Press d10) Display shows STRD stored message momentarily and then
advances to L.PNT Linearization Points Submenu.
If ENBL Known Loads scaling method was selected, calculate
the input values to the instrument based on the actual signal
being received. If DSBL Without Known Loads scaling method
was selected, calculate input values to the instrument based on
the transducer specification.
LINEARIZATION POINTS SUBMENU:
Press d11) Display flashes previous selection of Linearization Points
Submenu.
Press b12) Scroll though the available selections: 0002, 0003, 0004,
0006, 0006, 0007, 0008 , 0009, 0010 - up to 10
Linearization Points can be selected. Default is 0002.
If display flashes NONE, your instrument has only 2 linearization
points.
Press d13) Display shows STRD stored message momentarily only, if a
change was made, otherwise press ato advance to the FLTR
Filter Constant Submenu.
Linearization Points allow users to customize the Transducer curve.