34
ALARM 2 ENABLE/DISABLE SUBMENU:
Press d5) Display flashes previous selection. Press buntil ENBL
displays to use Alarm 2.
Press d6) Display shows STRD stored message momentarily and then
advances to ABSo only if it was changed, otherwise press ato
advance to ABSo Absolute/Deviation Submenu.
If DSBL Alarm 2 Disabled was selected, all submenus of Alarm
2 will be skipped and meter advances to LOOP Loop Break Time
Menu. If ENBL Alarm 2 Enabled was selected, Output 2 will
automatically be Disabled, and reassigned as Alarm 2.
The remaining Alarm 2 menu items are identical to Alarm 1 Menu.
Modifying Alarm Settings will not reset the instrument.
3.2.10 Loop Break Time MenuFigure 3.8 Flow Chart for Loop Break Time Menu
ENTER LOOP BREAK TIME MENU:
Press a1) Press a, if necessary, until CNFG prompt appears.
Press d2) Display advances to INPT Input Menu.
Press a3) Press a, if necessary, until Display advances to LOOP Loop
Break Time Menu.
It is required that you put the controller in the Standby Mode
for any configuration change other than Set Points & Alarms.