To minimize the possibility of overpressure damage, multi-range Note 9816 modules are manufactured to make use of the CAL (x) port of

the LOWEST installed pressure range.

4.3.2.Span Calibration Summary

Following is a simple, step-by-step procedure for executing a “full scale” span calibration of a 9816 Intelligent Pressure Scanner. It is assumed that all channels in the unit have the same full-scale pressure range. Optional commands are shown within brackets [ ].

Description

TCP/IP Data

 

Insure that valves in RUN/CAL mode (default).

[w1200]

 

 

 

 

 

 

… normal data acquisition

 

 

 

 

 

 

 

Perform Re-zero calibration

see Sec. 4.2.2

 

 

Place the module calibration manifolds into the CAL

[w0C01] for CAL pos.

 

 

position if this is the desired pressure application input.

[w0C00] for RUN pos.

 

 

The RUN position may be a better choice for modules with

 

 

transducers having different ranges.

 

 

 

Apply exact full scale pressure to the appropriate module

 

 

 

CAL and CAL REF inputs [or optionally to RUN inputs]

 

 

 

Delay for settling of pneumatic inputs

 

 

 

Verify that measured data reads near expected full scale

[rFFFF0]

 

 

 

 

 

 

Instruct module to calculate new gain coefficients for all 16

ZFFFF

 

 

channels

 

 

 

Place calibration manifold back into the RUN position

w0C00

 

 

Store new gain coefficients into transducer nonvolatile

w09

 

 

memory

 

 

 

… continue normal data acquisition

 

 

 

 

 

 