Pressure Systems 98RK-1 user manual Example, “v00800-01 0.000 1.000” Response returned is A

Models: 98RK-1

1 153
Download 153 pages 51.42 Kb
Page 93
Image 93
●Send TCP/IP command to 9816 module (i.e., via its connected socket) with replacement values for the channel’s offset and gain correction terms loaded into the module’s volatile memory (cc = 00-01). Load these into channel # 8's Transducer Coefficient array (aa=08):

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

Example:

Send TCP/IP command to 9816 module (i.e., via its connected socket) with replacement values for the channel’s offset and gain correction terms loaded into the module’s volatile memory (cc = 00-01). Load these into channel # 8's Transducer Coefficient array (aa=08):

v00800-01 0.000 1.000” Response returned is:

"A"

Send command to 9816 module (via its connected socket) to change its default EU output from psi to kPa. This will be done by changing the EU Pressure Conversion Scaler to 6.894757.

"V01101 6.894757" Response returned is

"A"