Pressure Systems 98RK-1 user manual LEAK Mode Inputs

Models: 98RK-1

1 153
Download 153 pages 51.42 Kb
Page 27
Image 27
2.3.4.5LEAK Mode Inputs

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

2.3.4.5LEAK Mode Inputs

The purge/leak-charge valve design includes a leak check feature capable of testing the integrity of user pneumatic connections as well as those within the 9816 module. For the leak mode to be used, all RUN mode pressure inputs must be dead-ended (closed) by the user. (Contact the PSI Sales Department for availability of an external leak check valve) When the 9816 is commanded into the LEAK/CHARGE position, the CAL 1-8input ports will be pneumatically connected to the module run side inputs. Common reference modules will connect only the numbered run side inputs to CAL (RUN REF is not charged). True differential (reference per port) modules will connect both the measurement input and reference port to CAL. While in the LEAK position, a test pressure may be applied thought the CAL port which will charge the dead-ended run side tubulation.

Note Test pressures applied to the CAL port during leak check operation must not exceed the full scale pressure of any internal transducer.

Once the lines are charged, the 9816 may be commanded back to the RUN position. This will reattach the charged run side lines to their corresponding internal transducer. Consecutive pressure readings from the 9816 will now allow user to determine the line leak rates. (Factory testing allows a maximum leak rate of 2% in two (2) minutes.) Once returned to the RUN position, lack of pressure indicates a gross leak. A slowly declining pressure indicates a slight leak. A leak is more difficult to detect as tubing volume increases. In the case of true differential units where both sides of the sensor are pressurized with the leak test pressure, an initial differential pressure of 0.0 psi should be measured when the unit is placed in the RUN position. If the measurement or RUN side of the channel leaks at a rate greater than the reference side, a resulting negative differential pressure will be measured. Likewise, if the reference port tubing leaks at a rate greater than the measurement side, a resulting positive differential pressure will be measured.

Figure 2.7

Page 25

www.PressureSystems.com

Page 27
Image 27
Pressure Systems 98RK-1 user manual LEAK Mode Inputs