5.1.6.4Calibration Manifold Piston O-Ring Replacement
Following is a
(1)Disassemble the module as described in Section 5.1.2.
(2)To minimize the possibility of damage, remove the
(3)Remove the tubing plate as described in Section 5.1.6.2.
(4)Using your index finger, shift the calibration manifold back and forth several times to loosen its connection with the adapter plate
(5)To remove the six (6) pistons from their slots on the manifold valve, hold the valve in one hand, and apply air pressure of approximately 30 psi (200 kPa) to the C3/C4 input passages on the valve. The passages (0.043" diameter holes) are located on the side of the valve next to the tubing plate, one hole on each side of the valve. Pressure on one side will release three (3) pistons, and pressure on the other side will release the other three (3). Thoroughly clean the pistons with a fast evaporating cleaning fluid that leaves little or no residue (e.g., acetone, alcohol, Freon, etc.) and dry with supply air. Replace the piston
(6)To remove the two (2) C1/C2 pistons from their slots on either end of the manifold valve housing, hold the manifold valve housing in one hand, and apply approximately 30 psi (200 kPa) to the two bulge tubes, one on either end of the housing
(7)Thoroughly clean the calibration manifold with a fast evaporating cleaning fluid that leaves little or no residue (e.g., acetone, alcohol, or Freon). Replace the calibration manifold into the housing, making sure that the guiding pin fits into the slot of the manifold housing.
(8)Replace the tubing plate as described in Section 5.1.6.2.
(9)Replace the
(10)Test your scanner to ensure proper operation.