●Send TCP/IP command to a 9816 module (via its open socket) to calculate and set gain coefficients for channels 8 through 4. Instruct the module to use 14.8890 psi as the applied pressure instead of each transducer’s full-scale value:

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

Note

Example:

The calculated gain values from the latest 'Z' command will be lost when the module is powered OFF. To save these gain terms to each transducer's nonvolatile memory, refer to the Set Operating Options ('w') command (index 09).

Send TCP/IP command to a 9816 module (via its open socket) to calculate and set gain coefficients for channels 8 through 4. Instruct the module to use 14.8890 psi as the applied pressure instead of each transducer’s full-scale value:

Z00F8 14.8890” Response:

" 1.01289 1.06953 1.03750 0.99704 1.10051"

Actual gain values are returned in the above response as decimal ASCII strings, each preceded by a space character. From left-to-right: they are for channels 8, 7, 6, 5, and 4.