Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

Example:

Send TCP/IP command to 9816 module (via its connected socket) that returns decimal raw “temperature” A/D counts data for channels 1, 5, 9, and 13:

m11110

Response contains data for channels 13, 9, 5, and 1 (left to right):

" 32767.000000 -32700.000000 10.000000 16385.000000"

Please note that channel 13 is saturated at +full scale, channel 9 is almost saturated at – full scale. Channel 5 reads near zero, and Channel 1 is about ½ +full-scale.

Page 77

www.PressureSystems.com

Page 79
Image 79
Pressure Systems 98RK-1 user manual M11110, 32767.000000 -32700.000000 10.000000