Read responses:

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

NOTE: With the exception of binary format 7, all other formats include a leading space in each datum delivered in each stream packet.

The number of stream packets (num) parameter is a positive integer count (specified with 1 to 5 numeric digits as needed). It sets a finite limit on the number of packets delivered in the host data stream. The value 0 for this parameter requests “continuous” output packets for the defined host stream (unbounded).

Example:

Configure three (3) separate autonomous host delivery streams, and divide the module’s channels between them. Channels (1-4) must be delivered to host at a 10 Hz rate, channels 5-8 may be delivered at half that rate, while the remaining channels 9-16 are delivered at half the previous rate. All streams are generated continuously and synchronized with the internal clock at 100 msec., 200 msec., and 400 msec. periods, respectively. Data are requested in single precision binary IEEE float format (f = 7).

c 00 1 000F 1 100 7 0” “c 00 2 00F0 1 200 7 0” “c 00 3 FF00 1 400 7 0

Read responses:

A” “A” “A

To similarly acquire data at “relative” rates (1, 2, and 4) using a periodic hardware trigger (assumed to also cycle at 10 Hz rate), enter the commands:

c 00 1 000F 0 1 7 0” “c 00 2 00F0 0 2 7 0” “c 00 3 FF00 0 4 7 0

Read responses:

"A"

"A"

"A"