Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

Figure 2.5

Each 9816 is manufactured with a connection to one of the eight (8) CAL port ranges, based on the unit’s full scale pressure range. Assignment of 9816's CAL range is show in Table 2.1 (next page). In the case of a 9816 with multiple pressure ranges installed, the unit will use the CAL range of the lowest installed full scale pressure range.

CAL port assignments are based on a pressure range, NOT 98RK-1 Note slot assignments.

Table 2.1

Pneumatic Connector Assignment

98RK-1

Connection

Pressure Scanner Range Input

Cal 1

User defined

 

 

Cal 2

10" W.C.; 20" W.C.; ±1 psid, ±1.5 psid

 

 

Cal 3

±2.5; ±5 psid

Cal 4

±10; ±15 psid, ±20 psid, ±25 psid

Cal 5

±30; ±45; ±50 psid

Cal 6

±75 psid, ±100 psid

Cal 7

+150; 200; 250; 300 psid

 

 

Cal 8

500 psid, 600 psid, 650 psid, 750 psid, 850

 

psid

Page 23

www.PressureSystems.com

Page 25
Image 25
Pressure Systems user manual Pressure Systems, Inc 98RK-1 & 9816 User’s Manual