5.1.7Front Pull-out Slide Tray

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

(7)Replace the transducer, making sure the electrical connections are located on the outside (back) edge of the cubic design. Tighten the 4-40 standoff screw snugly (finger tight plus ¼ turn).

(8)Reattach the flex circuit, ensuring it is attached to the same pins from which it was removed, and that the pins are not bent. Replace the 3/32" Allen-head stabilizing screw into each of the transducer 4-40 standoffs.

(9)Ensure all back panel electrical connections are in place, and replace the back panel by sliding it into position and re-installing the eight (8) Phillips-head retaining screws.

(10)Test your 98RK-1 Scanner Interface Rack to ensure proper operation.

5.1.7Front Pull-out Slide Tray

The front pull-out slide tray contains the PC-363 Ethernet controller board and the 98RK-1power supply. Neither of these components is considered to be field-repairable. If it is determined that maintenance is required on either of these units, the entire pull-out tray should be returned to Pressure Systems for replacement.

Switches for setting Cluster and Rack (physical) addresses are located on the PC-363 board (Cluster address) and on the front panel (Rack address). See Chapter 3, “Read Module Status” (Command ‘q’) for more information regarding scanner physical addresses.