Pressure Systems, Inc.98RK-1 & 9816 User’s Manual©

Now that data have been collected for every point

C 02

originally specified, calculate and apply the new

 

coefficient data with a Calculate & Apply (‘02’) sub-

 

command of ‘C’. This also restores the module to using

 

its original averaging parameters that existed before the

 

first ‘C’ command.

 

Place calibration manifold back into the RUN position, if

[w0C00]

the CAL position was used.

 

Store new offset and gain coefficients into transducer

w08

non-volatile memory.

w09

… continue normal data acquisition.

 

 

 

4.5Coefficient Storage

The various calibration functions described in Sections 4.2 and 4.3 update the active offset and gain coefficients, respectively, in the module’s volatile main memory (RAM) only. These newer calibration coefficients will be lost when instrument power is turned off. The Set Operating Options ('w') command may be used to also store these coefficients back in each transducer’s nonvolatile memory. This command’s option index = 08 will store new offset coefficients, while its option index = 09 will store new gain coefficients.

A user may read (and should verify) any new offset and/or gain coefficients after performing each calibration adjustment command (i.e., by saving coefficient data returned in a command 'h' or 'Z' response), or the Read Internal Coefficients ('u') command may be used to read them any time after calibration adjustment commands have been performed (see coefficient indexes cc=00 and 01 for arrays aa=01 through 10). These “adjusted” coefficients may be verified, and then saved by storing them in each transducer’s non-volatile memory with the 'w' command described above. Alternately, they may be verified and stored on the host computer’s secondary storage, and later restored (if necessary) with the Download Internal Coefficients ('v') command (same array/coefficient indexes as 'u').

4.6Line Pressure Precautions

When operating 9816 pressure scanners at elevated line or reference pressures, care must be taken when any command is issued that may result in shifting of the calibration valve. The user must ensure that any valve shifts will not result in the internal DH200 transducers being exposed to pressure transients that may exceed the proof pressure rating of the transducer. This is especially important when operating at elevated reference pressures as a shift to the CAL position may result in a rapid pressure change if the CAL/CAL REF pressure varies greatly from the measurement reference pressure.

Page 109

www.PressureSystems.com

Page 111
Image 111
Pressure Systems 98RK-1 user manual Coefficient Storage, Line Pressure Precautions