Pressure Systems 98RK-1 user manual RESET Command B, Response, Example

Models: 98RK-1

1 153
Download 153 pages 51.42 Kb
Page 45
Image 45
RESET (Command 'B')

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

RESET (Command 'B')

Purpose: Instructs the module to reset internal operating parameters, and to set all internal control variables to their default “reset” state (see description below). The current TCP/IP socket connection will remain open. Execution after a power off/on cycle is optional (unnecessary).

Command

B

 

‘B’ is the command letter

 

 

Response

A

 

‘A’ is the acknowledge letter

 

 

Description: The module returns to the following “reset” states if this command is executed:

Re-zero correction (offset) terms are set to the last values stored in transducer memory.

Span correction (gain) terms are set to the last values stored in transducer memory.

Calibration Valve is set to the RUN Position.

Number of Samples for Data Averaging is set to last value stored in non- volatile memory (factory default = 8).

Any autonomous host data delivery streams defined by ‘c’ sub-commands are reset (undefined).

The Multi-Point Calibration function defined by ‘C’ sub-commands is reset (undefined) if in progress.

Example:

Send TCP/IP command to a Model 9816 module (via open socket) to reset defaults: “B

Read following response:

"A"

Page 43

www.PressureSystems.com

Page 45
Image 45
Pressure Systems 98RK-1 user manual RESET Command B, Response, Example