Pressure Systems 98RK-1 Integrated Multi-Point Calibration Adjustment, Description, TCP/IP Data

Models: 98RK-1

1 153
Download 153 pages 51.42 Kb
Page 108
Image 108
see Section. 4.2.2

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

Following is a simple, step-by-step procedure for executing a “specified-value” span calibration of a Model 9816 Intelligent Pressure Scanner. For the purposes of this example, it will be assumed that an upscale pressure of 14.9800 psi is available from a dead weight tester for the calibration of 15 psi internal transducers. All sixteen channels are 15 psi full scale.

Description

TCP/IP Data

 

Insure that valves in RUN/CAL mode (default).

[w1200]

 

 

 

 

 

 

… normal data acquisition

 

 

 

 

 

 

 

Perform Re-zero calibration first

see Section. 4.2.2

 

 

Place the module calibration manifolds into the CAL

[w0C01] for CAL pos.

 

 

position if this is the desired pressure application input.

[w0C00] for RUN pos.

 

 

The RUN position may be a better choice for modules with

 

 

transducers having different ranges.

 

 

 

With a deadweight tester, apply 14.9800 psi to the

 

 

 

appropriate module CAL and CAL REF inputs [or optionally

 

 

 

to RUN inputs]

 

 

 

Delay for settling of pneumatic inputs

 

 

 

 

 

 

 

Verify measured data reads near expected full scale.

[rFFFF0]

 

 

 

 

 

 

Instruct module to calculate new gain coefficients for all 16

ZFFFF 14.98

 

 

channels

 

 

 

Place calibration manifold back into the RUN position

w0C00

 

 

 

 

 

 

Store new gain coefficients to transducer nonvolatile

w09

 

 

memory

 

 

 

… continue normal data acquisition

 

 

 

 

 

 

4.4Integrated Multi-Point Calibration Adjustment

NetScannerSystem Intelligent Pressure Scanners may have their accuracy enhanced by regular application of one or both of the individual Re-zeroand Span calibration adjustment functions already described in Sections 4.2 and 4.3. For users who have installed scanner firmware version 2.24 or later, an integrated and more comprehensive Multi-Point Calibration function may be used instead to adjust the same offset and gain coefficients for a module’s channels. This function also provides for supplying additional calibration points, which is particularly useful when it is desired to fit the adjustment data to the entire negative and positive range of the module. This integrated calibration adjustment function is fully described in this section. It is implemented by four (4) sub-commands of the 'C' command. Since this function combines the functions of the Re-zero and Span calibration adjustments (using the 'h' and 'Z' commands) it is recommended that you read the information of Section 4.2 and 4.3 before attempting to perform this multi-point calibration.

It is recommended that a Multi-Point Calibration adjustment be performed whenever new transducers are installed in your module. In such cases, the new zero and gain coefficients that

Page 106

www.PressureSystems.com

Page 108
Image 108
Pressure Systems 98RK-1 user manual Integrated Multi-Point Calibration Adjustment, Description, TCP/IP Data, see Section