●Configure and start the Multi-Point Calibration function so that it affects only the first four (4) channels of the module. Three (3) pressure calibration points will by supplied when we continue this function later (see example for ‘01’ sub-command below). A linear (1st order) fit will be used to obtain a new set of offset and gain correction coefficients for these four (4) channels. The maximum average sample count (64) is used to collect each calibration data point, so as to minimize any noise in the data samples. The module’s data acquisition process is altered immediately to collect the increased number of averages.

Pressure Systems, Inc.

98RK-1 & 9816 User’s Manual©

 

 

Example:

Configure and start the Multi-Point Calibration function so that it affects only the first four (4) channels of the module. Three (3) pressure calibration points will by supplied when we continue this function later (see example for ‘01’ sub-command below). A linear (1st order) fit will be used to obtain a new set of offset and gain correction coefficients for these four (4) channels. The maximum average sample count (64) is used to collect each calibration data point, so as to minimize any noise in the data samples. The module’s data acquisition process is altered immediately to collect the increased number of averages.

C 00 F 3 1 64” Read response:

"A"