BS-3680 U-Flume, P3 = 4

 

 

 

‘U’ parameters required *

‘U’ parameters calculated **

U0

= approach diameter

Da

U4

= Cv

U1

= throat diameter

D

U5

= Cd

U2

= hump height

p

U6

= Cu

U3

= throat length

L

U7

= A

*obtain from manufacturer’s specifications.

**calculated by OCM-3. May be viewed by accessing ‘U’ parameter.

Reference

ABSOLUTE CALCULATION, P4 = 0¤

 

For flows that can be calculated by the equation:

 

q = (2/3)1.5 x g0.5 x Cv x Cu x Cd x D x h1.5

 

where : q = flow rate

Cv = velocity coefficient

g = gravitational acceleration

Cu = shape coefficient

b = throat width

Cd = discharge coefficient

h = head

 

RATIOMETRIC CALCULATION, P4 = 1¤

 

For flows that can be calculated by the equation:

 

q = qcal x Cu/Cucal x Cd/Cdcal x Cv/Cvcal x (h/hcal)1.5

where : q = flow rate

qcal = flow rate at maximum head h = head

hcal = maximum head

Cu = shape coefficient for head

Cucal = shape coefficient for maximum head Cv = velocity coefficient

Cvcal = velocity coefficient for maximum head Cd = discharge coefficient for head

Cdcal = discharge coefficient for maximum head

¤Refer to Operation \ Flow Calculation.