Siemens 7ML19985AB01 instruction manual For h ≥ d, = B b/d Where q = flow rate, l/s

Models: 7ML19985AB01

1 112
Download 112 pages 47.88 Kb
Page 85
Image 85

MODIFIED TRAPEZOIDAL AREA X VELOCITY, P3 = 15

‘U’ parameters listed *

 

‘U’ parameters calculated **

U0

= channel top width

B

U3 = area (h)

U1

= channel base width

b

 

U2

= transition height

ht

 

*obtain from manufacturer’s specifications.

**calculated by OCM-3. May be viewed by accessing ‘U’ parameter.

Reference

ABSOLUTE CALCULATION, P4 = 0¤

For flows that can be calculated by the equation:

q = 1/1000 x (b + mh) x v for h d

 

q = 1/1000 x ((b + md)d + (B(h - d))) x v

for h d

m = (B - b)/d

 

where : q = flow rate, l/s

d = height of transition, cm

B = channel top width, cm

h = head, cm

b = channel base width, cm

v = flow velocity, cm/sec

RATIOMETRIC CALCULATION, P4 = 1¤

 

For flows that can be calculated by the equation:

 

q = qcal x A/Acal x v/vcal

 

for h d: A = (b + mh) x h

 

Acal = (b + mhcal) x hcal

 

for h d: A = (b + md) x d + B(h - d)

 

Acal = (b + md) x d + B(hcal - d)

 

m = (B - b)/d

 

where : q = flow rate

d = depth of channel

qcal = flow rate at hcal

h = head

B = channel top width, cm

v = flow velocity, cm/sec

b = channel base width, cm

 

¤Refer to Operation \ Flow Calculation.

7ML19985AB01

OCM III

86

Page 85
Image 85
Siemens 7ML19985AB01 instruction manual For h ≥ d, = B b/d Where q = flow rate, l/s