Siemens Installation Guide for the Milltronics OCM III Transducer and Temperature Sensor

Models: 7ML19985AB01

1 112
Download 112 pages 47.88 Kb
Page 16
Image 16

INSTALLING THE TRANSDUCER

Max cable run 183 m (600 ft) of RG-62U or equivalent. Cable must be run in a grounded metal conduit with no other cabling (except Temp. Sensor cable).

Ground shield at OCM-3 only.

Insulate shield at junctions to prevent inadvertent grounding.

Basic Wiring – Transducer

Hazardous voltage present on transducer terminals during operation.

Note: When using the XRS-5 transducer with the OCM-3, use the TS-2 external temperature sensor. The internal temperature sensor in the XRS-5 cannot be used.

INSTALLING THE TEMPERATURE SENSOR

In order to compensate for uniform temperature change in the air between the transducer and the flow surface, the temperature sensor must be connected to the OCM-3.

Maximum cable run 183 m (600 ft) of Beldon 8760, 1 pair shielded/twisted, 18 AWG or equivalent.

Temperature sensor cable can be run with the transducer cable in a grounded metal conduit.

Ground shield at OCM-3 only.

Basic Wiring – Temperature Sensor

16

OCM III

7ML19985AB01

Page 16
Image 16
Siemens 7ML19985AB01 Installing the Transducer, Installing the Temperature Sensor, Ground shield at OCM-3 only