INSTALLING THE TRANSDUCER
Max cable run 183 m (600 ft) of
Ground shield at OCM-3 only.
Insulate shield at junctions to prevent inadvertent grounding.
Basic Wiring – Transducer
Hazardous voltage present on transducer terminals during operation.
Note: When using the
INSTALLING THE TEMPERATURE SENSOR
In order to compensate for uniform temperature change in the air between the transducer and the flow surface, the temperature sensor must be connected to the
Maximum cable run 183 m (600 ft) of Beldon 8760, 1 pair shielded/twisted, 18 AWG or equivalent.
Temperature sensor cable can be run with the transducer cable in a grounded metal conduit.
Ground shield at OCM-3 only.
Basic Wiring – Temperature Sensor
16 | OCM III | 7ML19985AB01 |